Skip to main content

SAN++

Continuous flow analyzer

Skalar SAN++ Advanced analyzer with 1074 autosampler
SAN++ Advanced · The instrument shown may vary depending on your analyses.

From sampling
to measurement.

The SAN++ integrates every stage of continuous flow analysis. Its modules let you add the sample preparation each method requires.

Analytical principle
Segmented continuous flow
System design
Modular design
Sample preparation
In-line dialysis, digestion and distillation, depending on the method

Configurations

One platform.
Three configurations.

See the differences between Compact, Classic and Advanced.

SAN++

Compact

Skalar SAN++ Compact

Automation in less space.

Up to 3 analytical modules

Explore Compact

Analyzes one to three parameters at the same time. It can be expanded with additional racks for analytical modules.

Capacity
Up to 3 analytical modules
Peristaltic pump
1 × 21 pump tube positions
Detector space
2 dual channel detectors or 3 matrix detectors
Instrument dimensions
335 × 340 × 837 mm (H × D × W)
Compatibility
Compatible with previous-generation San+ and SAN++ analytical modules, and with the 1100 / 1150, 1074 and 1075 autosamplers.

The total space required also depends on the autosampler. Options for diluting samples and preparing calibration solutions vary with the model chosen. Automatic pump opening is optional.

SAN++

Classic

Skalar SAN++ Classic

Several analyses in one instrument.

Up to 5 analytical modules

Explore Classic

Analyzes up to five parameters at the same time. Components are chosen to match the analyses your lab needs.

Capacity
Up to 5 analytical modules
Peristaltic pump
2 × 16 pump tube positions
Detector space
4 dual channel detectors or 5 matrix detectors
Instrument dimensions
335 × 510 × 837 mm (H × D × W)
Compatibility
Compatible with previous-generation San+ and SAN++ analytical modules, and with the 1100 / 1150, 1074 and 1075 autosamplers.

The total space required also depends on the autosampler. Options for diluting samples and preparing calibration solutions vary with the model chosen.

SAN++

Advanced

Skalar SAN++ Advanced

More room for your analyses.

Up to 6 analytical modules

Explore Advanced

Holds more modules, including those that need more complex sample preparation. It can be expanded with additional racks.

Capacity
Up to 6 analytical modules
Peristaltic pump
2 × 21 pump tube positions
Detector space
4 dual channel detectors or 6 matrix detectors
Instrument dimensions
335 × 680 × 837 mm (H × D × W)
Compatibility
Compatible with previous-generation San+ and SAN++ analytical modules, and with the 1100 / 1150, 1074 and 1075 autosamplers.

The total space required also depends on the autosampler. Options for diluting samples and preparing calibration solutions vary with the model chosen. Automatic pump opening is optional.

Each module performs one part of the analysis. The number of parameters you can measure depends on how the instrument is configured.

In all three models

The capacity changes.
The foundation stays the same.

Control at your fingertips
Touch screen and FlowAccess V3 to control the instrument.
Fewer manual tasks
Schedule start-up, shut-down and rinse cycles.
Integrated preparation
Add the options your method requires.

Technology

Four stages.
One integrated system.

See how sampling, preparation and measurement connect, all the way to the results.

01 / Sampling

Skalar 1074 autosampler.
Skalar 1074 autosampler.
01SamplingHow your samples enter the system.

The autosampler feeds the samples into the system. The choice depends on how many you process and the containers you use. Depending on the configuration, it can also dilute samples and prepare calibration solutions.

1100 / 1150
2 × 50 sample positions
1074
Up to 300 samples
1075
576 sample positions + 26 for standards and controls

Check container compatibility, dilution and number of needles before choosing the sampler.

02PreparationThe treatments your method needs.

Preparation is built into the sample’s path. Dialysis, digestion and distillation options are chosen according to what your method requires.

Dialysis
In-line treatment depending on the application
UV digestion
Integrated in the chemistry module
Distillation
Preparation within the analytical path
Throughput
30–120 analyses/hour, depending on the application

Throughput is not a fixed capacity for every method and does not necessarily equal samples per hour.

03MeasurementHow we measure what you want to analyze.

The detector is chosen based on what you need to measure. Options include colorimetry, UV, fluorimetry and ion-selective electrodes.

Colorimetry
Digital dual channel detectors
Matrix correction
Option for specific applications
Other technologies
UV · Fluorimetry · ISE

The choice depends on the substance you need to measure, its concentration and the sample type.

04ResultsInstrument control, monitoring and reports.

With FlowAccess you can control the instrument, follow measurements and organize results and reports.

Planning
Sample tables and dilutions
Monitoring
Real-time signals and results
Quality control
Configurable CLP / GLP criteria
FlowReport
Custom reports and export

System options are chosen based on your analyses, the method and the model.

Applications

What do you need to analyze?

Explore SAN++ applications by sample type and the parameters you want to measure.

Water and environment

Nutrients and other parameters for water analysis. Preparation options are defined by the sample and the method.

Key parameters

  • Ammonia
  • Nitrate + nitrite
  • Nitrite
  • Ortho-phosphate
  • Total phosphorus
  • Total nitrogen
  • Chloride
  • Sulfate
  • Silicate
  • Total cyanide
  • Phenol index
  • Anionic surfactants (MBAS)
Methods and configuration

Preparation

Integrates digestion, distillation, dialysis or gas diffusion depending on the determination. The treatment is matched to the sample and the analytical method.

Analytical methods

Options based on EPA 350.1 for ammonia, EPA 365.1 for phosphate and Kelada 01 for total cyanide. Each method requires its own configuration and procedure.

Beer and malt

Analyses to monitor brewing and evaluate malt, with dedicated modules for each parameter.

Key parameters

  • Bitterness
  • Free amino nitrogen (FAN)
  • Diacetyl
  • β-glucans
  • Diastatic power
  • α-amylase
  • Polyphenols
  • Sulfur dioxide
  • Color
  • Total reducing sugars
Methods and configuration

Applications

Combines bitterness, total SO₂, FAN, polyphenols and diacetyl for beer. For malt, it integrates diastatic power, α-amylase, FAN, β-glucans and color.

Methods and detection

Methods based on EBC, ASBC and MEBAK, with UV/VIS or fluorimetric detection depending on the parameter and configuration.

Wine

Determinations to control winemaking and wine composition.

Key parameters

  • Free SO₂
  • Total SO₂
  • Volatile acidity
  • Total acidity
  • Malic acid
  • Lactic acid
  • Acetic acid
  • Glucose / fructose
  • Total reducing sugars
  • Ethanol
Methods and configuration

Applications

Combines free and total SO₂, volatile acidity, malic acid and total acidity. Each module is configured for a specific determination.

Preparation

In-line dialysis handles color interference. Automatic standard preparation and sample dilution are added as the application requires.

Soil, plant and fertilizer

Nutrient analysis in extracts and digests, with configurations suited to how each sample is prepared.

Key parameters

  • Ammonia nitrogen
  • Nitrate + nitrite
  • Phosphorus
  • Potassium
  • Boron
  • Calcium
  • Magnesium
  • Urea
  • Zinc
  • Sulfate
Methods and configuration

Preparation

Determines ammonia nitrogen in extracts and nitrogen after Kjeldahl digestion. For phosphorus, it works with Bray or Olsen extracts and digests, using the preparation each method calls for.

Configuration

NPK configurations combine nitrogen, phosphorus and potassium. They can add boron, urea, zinc or sulfate depending on the lab’s analyses.

Tobacco

Determination of tobacco components with dedicated analytical modules.

Key parameters

  • Total alkaloids / nicotine
  • Sugars
  • Chloride
  • Ammonia
  • Nitrate + nitrite
  • Nitrogen
  • Phosphate
  • Cyanide
Methods and configuration

Applications

Combines nicotine, total sugars, chloride, ammonia and nitrate in a configuration suited to your analyses.

Methods and preparation

Methods based on ISO, CORESTA and AOAC, with in-line dialysis options to handle color interference depending on the application.

The configuration depends on the sample type, the parameter and the analytical method.

Frequently asked questions

Before you choose
your SAN++.

What do I need to define to choose a configuration?

The sample type, the parameters you need to measure, their expected concentrations and the analytical methods. Daily workload and available space complete the choice of modules and autosampler.

Do all analyses require the same sample preparation?

No. Each method sets its own preparation. The system can integrate in-line treatments such as dialysis, digestion and distillation; other applications require extracts or digests prepared before the sample is introduced.

Can I analyze different parameters at the same time?

Yes. The modular design lets you combine simultaneous determinations. The number of parameters depends on the model, the installed modules and the preparation each method requires; it is not automatically the same as the number of modules.

What determines how many samples I can process per day?

The method, preparation, dilutions and the sequence of calibrations and controls all affect throughput. Autosampler capacity shows how many samples it can hold, not how many are analyzed per hour.

Can I expand the system if my needs change?

The system’s configuration can be adapted. Compact and Advanced accept additional racks for analytical modules. Expansion depends on space for modules and detectors, pump capacity and the requirements of the new methods.

What is the difference between controlling the instrument and managing the results?

The touch screen controls basic functions, such as pump speed and reactor temperature. FlowAccess coordinates the analysis, displays the signals and calculates results. FlowReport lets you prepare and export reports.

Can it dilute samples and prepare standards automatically?

Yes, with the diluter station and the matching sampling options. The system can prepare working standards, dilute samples before analysis and re-analyze those above the calibration range. We help you define these functions for your methods.

The SAN++ in operation

Your next SAN++

Let’s define your
configuration.

Tell us what you need to analyze. We’ll help you define the right instrument, sample preparation and options for your lab.

You can also write to us

contacto@ugs-international.com.mx

Fields marked * are required.

Your information will be used to handle and follow up on your request. Privacy notice in Spanish (opens in a new tab).

How we handle your data

U.G.S. INTERNATIONAL S.A. DE C.V., located at Sierra Tarahumara 207, Lomas de Chapultepec, 11000 Mexico City, will use your identification and contact data, the information in your request and technical data from the submission to respond to you, prepare the service or quote you requested, follow up with you and protect the form. We do not request sensitive personal data, and sending an inquiry does not subscribe you to advertising. You may limit the use or disclosure of your data and exercise your rights by writing to contacto@ugs-international.com.mx.